IMPAX 6.5.1 Client Knowledge Base: Extended > Administering IMPAX > Configuring image display > Setting hanging protocols > Configuring ultrasound multistage stress protocols
When authenticating with a smart card, impersonation is not available.
Ultrasound multistage stress protocols must be enabled before the hanging protocol can be created.
To enable ultrasound multistage stress protocols
If setting preferences for other users, impersonate the user or role.
Hanging protocols created at the role level can be edited or removed only when impersonating that role. They are visible to but not editable by any lower-level users or roles. They are not visible to any higher-level roles. Hanging protocols created at the user level can be viewed, edited, or removed only by the user who created them; they are not visible to higher-level roles.
Press F12 or, from the Image area top toolbar, click Customize. ![]()
In the Customize dialog, switch to the Wizards tab.
Ensure that the Echo US Protocol wizard is listed and is enabled. If not, select it and click Enable.
To save the changes and close the Customize dialog, click OK.
Topic number: 49129 Applies to: IMPAX 6.5.1 Client Knowledge Base |