IMPAX 6.5.1 Server Knowledge Base home > Configuring IMPAX Server with Administration Tools > Study Presentation Manager

Display function settings: Reference


Use these options when creating or modifying display response functions.

Video or p-value Data

Options How it is used by IMPAX
Data are Linear p-values After the window level, pixel values are mapped directly to p-values as defined by DICOM. This is a perceptually linear response shape based on the Barton method of human visual response.

Gamma-Corrected Video

Pixel values are intended to be displayed on an output device with a gamma response such as a video monitor.

This includes video capture data and most older modalities.